ATOMICABSORPTIONSPECTROPHOTOMETRYCOOKBOOK
Section4
MeasuringConditionsbyElementofFurnaceAnalysisMethod
AtomicAbsorptionSpectrophotometryCookbook
Section4
CONTENTS
7.MeasuringConditionsbyElementofFurnaceAnalysisMethod1
7.1Measuringconditions1
7.2Applicationtoactualsamples2
7.3Typesandlifeofgraphitetube2
7.4Interferenceandbackgroundabsorptionandmeasuringcautions3
7.5Measuringconditionsbyelement5
1)Silver(Ag)6
2)Aluminum(Al)8
3)Arsenic(As)10
4)Gold(Au)12
5)Beryllium(Be)14
6)Bismuth(Bi)16
7)Calcium(Ca)18
8)Cadmium(Cd)20
9)Chromium(Cr)22
10)Cobalt(Co)24
11)Copper(Cu)26
12)Iron(Fe)28
13)Potassium(K)30
14)Lithium(Li)32
15)Magnesium(Mg)34
16)Manganese(Mn)36
17)Molybdenum(Mo)38
18)Sodium(Na)40
19)Nickel(Ni)42
20)Lead(Pb)44
21)Platinum(Pt)46
22)Rhodium(Rh)48
23)Antimony(Sb)50
24)Selenium(Se)52
25)Silicon(Si)54
26)Tin(Sn)I56
27)Tin(Sn)II58
28)Strontium(Sr)60
29)Titanium(Ti)62
30)Vanadium(V)64
31)Zinc(Zn)66
Measuringconditionsbyelementoffurnaceanalysismethod7.1Measuringconditions
Thesolutionpreparedonthebasisofthestandardsamplepreparationmethod(cookbooksection2-3)wasusedtomakethecalibrationcurveinthestandardconcentrationrange.AnalysisconditionsstoredinAA-6401or6501weremainlyused.HCL,BGC-D2,andBGC-SRmethodswereusedaccordingtotherespectiveelement.Therelationshipbetweentheashingtemperature,theatomizationtemperature,andsensitivityisalsoshown.However,theyareomittedforsomeelements.
Accordingtothedata,aStepheatingstagecomesaftertheRampheatingstage.BetterrepeatabilitycanbeexpectedbyinsertinganotherRampheatingstageuptotheAshingtemperaturebetweenthetwo.AcleaningstageissetaftertheAtomizationstageitneccessary.
Thehighdensitygraphitetubeorthepyrolyticgraphitetubewasuseddependingonthecharacteristicofeachelementanditsselectioniswrittenineachdata.
Astheatomizationcondition,therearetwomodes:thestandardmode(0.0)andthehighsensitivitymode(0.0H)areavailableandtwotypesofdataareshown.
Thestandardmodeandhighsensitivitymodeareusedasshownbelow.
Standardmode(0.0)
Theabsorbancesensitivityis1/3to1/2ofthatobtainedinthehighsensitivitymode,dependingon
theelement.
However,thismodeiseffectiveforthemeasurementofthesamplewhichissusceptibletoinfluencesfromco-existentsubstances.Backgroundabsorbanceisalsosmall.
Highsensitivitymode(0.0H)
Thehighestsensitivitycanbeobtainedinthismodebut,atthesametime,backgroundishigher.
Thisisparticularlyeffectiveforthemeasurementoftraceelementsincitywater,ultra-purewaterandothersampleswhichcontainsmallco-existentsubstances.
However,thestandardmodeisbetterformeasurementofthesamplewhicheasilycausesbackgroundabsorption.
Thisdataismeasuredattheinnergasflowof0l/minintheatomizationmode.Absorbancesensitivitycanbecontrolledbyflowingtheinnergas(0to1.5l/min)inthisstep.Whentheinnergasflowis1.5l/min,absorbancesensitivitydropstoabout1/5andthehighconcentrationcalibrationcurvecanbeprepared.
7.2Applicationtoactualsamples
Thetemperatureprogramofthisdataisdesignedforthestandardsolution.Goodresultsmaynotbeobtainedintheheatingconditiondescribedhere,becausematrixcomponents,concentration,andthetypeandconcentrationofco-existentsubstancesaredifferentintheactualsample.
Therefore,whenamatrixcomponentiscontainedortheconcentrationsofco-existentsubstancesarehigh,thesuitableconditionsfortheactualsampleshouldbesetbythemethodoftheanalysisconditioninthecookbooksection2-2-2FurnaceAnalysisMethod,onthebasisoftheheatingconditionofthisdata.
7.3Typesandlifeofgraphitetube
Thefollowingthreetypesofgraphitetubesaresupplied,anduseddependingontheelementtobemeasuredandtheinterferencelevel.
Highdensitygraphitetube
Thiscanbeusedforthemeasurementofmostelementsandcanobtainexcellentresultsin
repeatabilityoftheabsorptionsignal.However,withB,Ca,Moandothermetalswhicheasilyreactoncarbon,thetargetcomponentdoesnotvolatilizecompletelyduringatomizationandmemoryeffectoccurs.Withanorganicsolventtypesample,penetrationintothegraphitetubeincreasesandsensitivitymaydropsharplyormemoryeffectmayoccur.
Pyrolyticgraphitetube
Asurfacetreatmentisdoneonthehighdensitygraphitetubetomakeapyrolytictube.Because
samplepenetrationintothetubeissmall,excellentresultscanbeobtainedwithanorganicsolventtypesample.ThememoryeffecthardlyoccurswithB,Ca,Mo,etc.mentionedabove.
Itsabsorbancesensitivityistwotothreetimeshigherthanthatofthehighdensitygraphitetubewith
AlandSiand4b,5b,6b,7b,8groups(Ti,V,Cr,Mo,Fe,Co,Ni,Cu,etc.)andlanthanoidsintheperiodictableofelements.
Itsrepeatabilityoftheabsorptionsignalmaybeworsethanthatofthehighdensitygraphitetube.
Platformtypegraphitetube
Agraphiteplateof1mmthicknessandseveralmm2sizeissetinthegraphitetubetomaketheplatformtypegraphitetube.Becausethistubeheatsthesampleindirectlyandevenly,interferenceofco-existentsubstancesandbackgroundabsorptiondecrease.Thisiseffectivefortheanalysisofmetalsamplescontainingmatrixcomponentsandelementsandinbloodorseawater;particularlyforanalysisofCd,Pbandothermetalswhichhavelowboilingpoints.
Thegraphitetubecanbeuseduntilitisbroken,unlessasensitivityorrepeatabilitydropoccurs.Thislifevarieswidelydependingonliquidpropertiesofthesampletobemeasured.Itcanbeused2000timeswiththesamplewhichhardlycontainsanyacidorco-existentsubstances.
However,itsusemaybelimitedto300timeswiththesamplewhichhashighacidconcentrationoralotofco-existentsubstances.
Therefore,itisrecommendedtocountservicetimesofthegraphitetubeinmeasurementandcheckthetubelifeforthesampleinadvance.TheAA-6401or6501hastheprogramtocountservicetimesofthetube,sothiscanbeusedtoknowthetimewhenthetubeshouldbechanged.
7.4Interference,backgroundabsorption,andmeasuringcautions
Cautionsformeasuringeachelementandinterferenceareenumeratedbelow.
Somesamplesmaynotbemeasuredproperlysimplybydevisingtheheatingprogram.Insuchcase,goodresultsmaybeobtainedifthematrixmodifierisaddedinaccordancewiththecookbooksection2-5.3(Chemicalinterferenceanditscorrectionmethod).
Whenbackgroundabsorptioncorrectionisdone,asensitivitydropoccursintheBGC-SRmethod,ascomparedwithBGC-D2methodfortheprinciplereason.Iftheelementbeinganalyzedisatthe430nmorlongerwavelengthrange,theBGC-SRmethodshouldbeusedforbettercorrectionaccuracy.
Al
Argasisusedasthegas.WhenNgasexists,strongmonocyanideisproduced.
2
As
SensitivityslightlyincreasesbyusingN2ratherthanArasthesheathgas.Theashingtemperature
canbeincreasedbyaddingMg(NO)
2
orNi(NO).
3
32
Ca
Becausethegraphitetubeproducesacarbide,useofthepyrolytic
graphitetubeispreferable.
Attentionmustbegiventotheambientatmosphere,theregent,anddistilledwaterbeingusedfor
microchemicalanalysis.
Cd
Ashingtemperaturecanbeincreasedbyaddingabout1%(NH)
HPO.
42
4
Co
Interferenceoccursifahalogenideco-exists.
3
2
2)
Thesensitizationeffectcanbeexpectedbyadding0.5%Fe(NO).
Cr
Argasispreferableasthesheathgas.
Ifahalogenideco-exists,usethepyrolytic
graphitetube.Chromiumvolatilizesevenatlow
temperaturesettings.ItbecomesCrCl3undertheco-existenceofNaCl.
Cu
Fe
Ga
K
3)
Interferenceoccursifahalogenideco-exists.Itispreferabletodecomposeitbywetashing,oraddnitricacid,beforethedryingorashingconditionisset.
Itisrelativelyinsusceptibletointerferencesfromotherelements.However,attentionmustbegiventoFeimpuritiesinthegraphitetubeandambientatmosphereduringmicrochemicalanalysis.
ItisrecommendedtouseHNO3totreatastandardorunknownsample.ThedoublesensitizationeffectcanbeexpectedwiththemixedsolutionofHNO3+H2O2.1)
Attentionmustbegiventodistilledwaterandtheambientatmosphereduringmicrochemicalanalysis.
Itisdesirabletonarrowtheslitwidthformeasurement,becauseemissionfromthegraphitetubeisstronginthiswavelengthrange.
Mg
Mn
Mo
Na
Ni
Pb
Se
Si
Sn
Ti
V
Zn
Argasispreferableasthesheathgas.Attentionmustbegiventodistilledwaterandimpuritiesinthereagent.
Itissusceptibletointerferenceifahologenideco-exist.Itisexpectedthattheashingtemperature
o
canbeincreasedfrom1100to1150CbyaddingHNO3.
Carbideisproducedbycarbonfromthegraphitetubeorthesample,andsensitivitydrops.
Therefore,usingthepyrolyticgraphitetubeisrecommended.
Attentionmustbegiventotheambientatmosphere,anddistilledwaterduringmicrochemicalanalysis.
o
Volatizationissmallevenifashingtemperatureissetfrom900to1000C.
o
Volatizationincreaseswhenashingtemperatureissetfrom600to650Corhigher.
o
Itisdesiredthattheashingtemperatureis300Corlower.Itisexpectedthattheashingtemperature
o
isincreasedfrom1000to1100CbyaddingCu,NiandMo.
TheuseofN2gasispreferabletoArasthesheathgas.
Usingthepyrolyticgraphitetubeisrecommended.
Arispreferableasthesheathgas,andhighersensitivitycanbeobtainedwithasmallerflow.
o
Sensitivityislowandtheatomizationtemperatureis2700to3000C.Thepyrolyticgraphitetubeisused.
o
Volatizationissmalleveniftheashingtemperatureissetat700C.Theuseofthepyrolyticgraphitetubeispreferable.
Attentionmustbegiventotheatmosphereanddistilledwaterduringmicrochemicalanalysis.ThesensitizationeffectcanbeobtainedbyaddingAl(NO3)2.4)
Reference
R.D.Ediger,At.Absorpt.Newsl.,14,127-130(1975)
AAnalysiswiththeGraphiteFurnaceUsingMatrixModification”.
R.B.CruzandJ.C.VanLoon,Anal.Chim.Acta,72,231-243(1974)
.CriticalstudyofGraphiteFurnaceNon-FlamlAAStotheDeterminationofTraceBaseMetalsin
ComplexHeavyMatrixSampleSolutions”.
W.J.Price,SpectrochemicalAnalysisbyAtomicAbsorption,HEYDEN.
F.J.M.Maesen,J.BalkeandR.Massee,Spectrochim.Acta,33B,311-324(1978)
on-SpectralInterferencesinFlamelessAASUsingGraphiteMiniTube-Furnaces”.
7.5Measuringconditionsbyelement
Silver(Ag)
Standardsubstanceandstandardsolutionpreparationmethod
RefertoSection3-3forthestandardsolutionpreparationmethod
Measuringconditionsandcalibrationcurve
Lightingconditions
Current
;10mA/0mA
;10mA/400mA
Wavelength
;328.1nm
;328.1nm
Slitwidth
;
0.5nm
;
0.5nm
Lightingmode
;
BGC-D2
;
BGC-SR
Graphitefurnacetemperatureprogram(Highsensitivitymode0.0H),Highdensitygraphitetube
STAGE#
TEMP.(oC)
TIME(sec)
GAS
(l/min)
HEAT
SAMPLING
PreTIME
1
120
20
#1
1.0
RAMP
OFF
2
250
10
#1
1.0
RAMP
OFF
3
400
10
#1
1.0
STEP
OFF
4
400
3
#1
0.0H
STEP
OFF
5
1500
3
#1
0.0H
STEP
ON
2
CalibrationcurveinBGC-D2mode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)328.1nm
10.00000.0039
22.00000.2014
36.00000.5895
410.00000.9301
[ABS]=K2*[C]2+K1*[C]+K0r2=0.9999
K0=0.0016,K1=0.1043,K2=–0.0011
CalibrationcurveinBGC-SRmode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)328.1nm
10.00000.0057
22.00000.0868
36.00000.2843
410.00000.4801
Silver(Ag)
[ABS]=K1*[C]+K0r2=0.9991
K0=0.0012,K1=0.0479
Graphitefurnacetemperatureprogram(Standardmode0.0),Highdensitygraphitetube
STAGE#
TEMP.(oC)
TIME(sec)
GAS
(l/min)
HEAT
SAMPLING
PreTIME
1
120
20
#1
1.0
RAMP
OFF
2
250
10
#1
1.0
RAMP
OFF
3
400
10
#1
1.0
STEP
OFF
4
400
3
#1
0.0
STEP
OFF
5
1500
3
#1
0.0
STEP
ON
2
CalibrationcurveinBGC-D2mode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)328.1nm
10.00000.0033
22.00000.1217
36.00000.3734
410.00000.6077
[ABS]=K2*[C]2+K1*[C]+K0r2=0.9999
K0=0.0010,K1=0.0629,K2=–0.0002
CalibrationcurveinBGC-SRmode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)328.1nm
10.00000.0062
22.00000.0572
36.00000.1828
410.00000.3060
[ABS]=K1*[C]+K0r2=0.9991
K0=0.0018,K1=0.0303
Ashingtemperature,atomizationtemperatureandsensitivity(Solution:HNO3,AgNO3)
AshingconditionAtomizationcondition
Aluminum(Al)
Standardsubstanceandstandardsolutionpreparationmethod
RefertoSection3-3forthestandardsolutionpreparationmethod
Measuringconditionsandcalibrationcurve
Lightingconditions
Current
;8mA/0mA
;10mA/600mA
Wavelength
;309.3nm
;309.3nm
Slitwidth
;
0.5nm
;
0.5nm
Lightingmode
;
BGC-D2
;
BGC-SR
Graphitefurnacetemperatureprogram(Highsensitivitymode0.0H),Pyrolyticgraphitetube
STAGE#
TEMP.(oC)
TIME(sec)
GAS
(l/min)
HEAT
SAMPLING
PreTIME
1
120
20
#1
1.0
RAMP
OFF
2
250
10
#1
1.0
RAMP
OFF
3
700
10
#1
1.0
STEP
OFF
4
700
3
#1
0.0H
STEP
OFF
5
2400
3
#1
0.0H
STEP
ON
2
CalibrationcurveinBGC-D2mode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)309.3nm
10.00000.0035
24.00000.0896
312.00000.2846
420.00000.4353
[ABS]=K2*[C]2+K1*[C]+K0r2=0.9989
K0=0.0010,K1=0.0255,K2=–0.0002
CalibrationcurveinBGC-SRmode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)309.3nm
10.00000.0028
24.00000.0381
312.00000.1298
420.00000.1950
[ABS]=K1*[C]+K0r2=0.9949
K0=0.0030,K1=0.0098
Graphitefurnacetemperatureprogram(Standardmode0.0),pyrolyticgraphitetube
STAGE#TEMP.(oC)TIME(sec)GAS(l/min)HEATSAMPLINGPreTIME
1
120
20
#1
1.0
RAMP
OFF
2
250
10
#1
1.0
RAMP
OFF
3
700
10
#1
1.0
STEP
OFF
4
700
3
#1
0.0
STEP
OFF
5
2400
3
#1
0.0
STEP
ON
2
?CalibrationcurveinBGC-D2mode(Sampleinjectionvolume20l)
STD.
CONC.
ABS.
No.
(ppb)
309.3nm
1
0.0000
0.0035
2
4.0000
0.0497
3
12.0000
0.1653
4
20.0000
0.2914
[ABS]=K1*[C]+K0r2=0.9975
K0=–0.0030,K1=0.0145
CalibrationcurveinBGC-SRmode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)309.3nm
10.00000.0037
24.00000.0219
312.00000.0802
420.00000.1399
[ABS]=K1*[C]+K0r2=0.9953
K0=0.0010,K1=0.0069
Ashingtemperature,atomizationtemperatureandsensitivity(Solution:HClMetal)
AshingconditionAtomizationcondition
Arsenic(As)
Standardsubstanceandstandardsolutionpreparationmethod
RefertoSection3-3forthestandardsolutionpreparationmethod
Measuringconditionsandcalibrationcurve
Measuringconditions
Current
;12mA/0mA
;12mA/500mA
Wavelength
;193.7nm
;193.7nm
Slitwidth
;
0.5nm
;
2.0nm
Lightingmode
;
BGC-D2
;
BGC-SR
Graphitefurnacetemperatureprogram(Highsensitivitymode0.0H),highdensitygraphitetube,Pd10
ppmadded
STAGE#
TEMP.(oC)
TIME(sec)
GAS
(l/min)
HEAT
SAMPLING
PreTIME
1
120
20
#1
1.0
RAMP
OFF
2
250
10
#1
1.0
RAMP
OFF
3
600
10
#1
1.0
STEP
OFF
4
600
3
#1
0.0H
STEP
OFF
5
2200
3
#1
0.0H
STEP
ON
2
CalibrationcurveinBGC-D2mode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)193.7nm
10.00000.0113
24.00000.0996
312.00000.2456
420.00000.3922
[ABS]=K1*[C]+K0r2=0.9989
K0=0.0174,K1=0.0189
CalibrationcurveinBGC-SRmode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)193.7nm
10.00000.0085
24.00000.0366
312.00000.0838
420.00000.1219
[ABS]=K1*[C]+K0r2=0.9945
K0=0.0119,K1=0.0056
?
Graphitefurnacetemperatureprogram(Standardmode0.0),Highdensitygraphitetube,Pd10ppmadded
STAGE#
TEMP.(oC)
TIME(sec)
GAS
(l/min)
HEAT
SAMPLING
PreTIME
1
120
20
#1
1.0
RAMP
OFF
2
250
10
#1
1.0
RAMP
OFF
3
600
10
#1
1.0
STEP
OFF
4
600
3
#1
0.0
STEP
OFF
5
2200
3
#1
0.0
STEP
ON
2
CalibrationcurveinBGC-D2mode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)93.7nm
10.00000.0158
24.00000.0569
312.00000.1534
420.00000.2412
[ABS]=K1*[C]+K0r2=0.9994
K0=0.0114,K1=0.0114
CalibrationcurveinBGC-SRmode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)193.7nm
10.00000.0120
24.00000.0243
312.00000.0476
420.00000.0750
[ABS]=K1*[C]+K0r2=0.9987
K0=0.0116,K1=0.0031
Ashingtemperature,atomizationtemperatureandsensitivity(Solution:NaOH,As2O3)
AshingconditionAtomizationcondition
Gold(Au)
Standardsubstanceandstandardsolutionpreparationmethod
RefertoSection3-3forthestandardsolutionpreparationmethod
Measuringconditionsandcalibrationcurve
Lightingconditions
Current
;12mA/0mA
;10mA/400mA
Wavelength
;242.8nm
;242.8nm
Slitwidth
;
0.5nm
;
0.5nm
Lightingmode
;
BGC-D2
;
BGC-SR
Graphitefurnacetemperatureprogram(Highsensitivitymode0.0H),Highdensitygraphitetube
STAGE#
TEMP.(oC)
TIME(sec)
GAS
(l/min)
HEAT
SAMPLING
PreTIME
1
120
20
#1
1.0
RAMP
OFF
2
250
10
#1
1.0
RAMP
OFF
3
500
10
#1
1.0
STEP
OFF
4
500
3
#1
0.0H
STEP
OFF
5
1700
3
#1
0.0H
STEP
ON
2
CalibrationcurveinBGC-D2mode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)242.8nm
10.00000.0023
210.00000.1088
330.00000.3908
450.00000.6341
[ABS]=K2*[C]2+K1*[C]+K0r2=0.9984
K0=–0.0056,K1=0.0129,K2=–0.0000
CalibrationcurveinBGC-SRmode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)242.8nm
10.00000.0061
210.00000.0796
330.00000.2228
450.00000.3609
[ABS]=K2*[C]2+K1*[C]+KOr2=1.0000
K0=0.0061,K1=0.0074,K2=–0.0000
Graphitefurnacetemperatureprogram(Standardmode0.0),Highdensitygraphitetube
STAGE#TEMP.(oC)TIME(sec)GAS(l/min)HEATSAMPLINGPreTIME
1
120
20
#1
1.0
RAMP
OFF
2
250
10
#1
1.0
RAMP
OFF
3
500
10
#1
1.0
STEP
OFF
4
500
3
#1
0.0
STEP
OFF
5
1700
3
#1
0.0
STEP
ON
2
?CalibrationcurveinBGC-D2mode(Sampleinjectionvolume20l)
STD.
CONC.
ABS.
No.
(ppb)
242.8nm
1
0.0000
0.0029
2
10.0000
0.0922
3
30.0000
0.2784
4
50.0000
0.4494
[ABS]=K2*[C]2+K1*[C]+K0r2=0.9999
K0=0.0013,K1=0.0095,K2=–0.0000
CalibrationcurveinBGC-SRmode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)242.8nm
10.00000.0073
210.00000.0600
330.00000.1820
450.00000.2959
[ABS]=K2*[C]2+K1*[C]+K0r2=0.9996
K0=0.0055,K1=0.0058,K2=0.000
Ashingtemperature,atomizationtemperatureandsensitivity(Solution:AquaRegia,Metal)
AshingconditionAtomizationcondition
Beryllium(Be)
Standardsubstanceandstandardsolutionpreparationmethod
RefertoSection3-3forthestandardsolutionpreparationmethod
Measuringconditionsandcalibrationcurve
Lightingconditions
Current
;16mA/0mA
Wavelength
;234.9nm
Slitwidth
;
0.5nm
Lightingmode
;
BGC-D2
Graphitefurnacetemperatureprogram(Highsensitivitymode0.0H),Pyrolyticgraphitetube
STAGE#
TEMP.(oC)
TIME(sec)
GAS
(l/min)
HEAT
SAMPLING
PreTIME
1
120
20
#1
1.0
RAMP
OFF
2
250
10
#1
1.0
RAMP
OFF
3
600
10
#1
1.0
STEP
OFF
4
600
3
#1
0.0H
STEP
OFF
5
2600
3
#1
0.0H
STEP
ON
2
?CalibrationcurveinBGC-D2mode(Sampleinjectionvolume20l)
STD.
CONC.
ABS.
No.
(ppb)
234.9nm
1
0.0000
0.0041
2
0.2000
0.1367
3
0.6000
0.4307
4
1.0000
0.7463
[ABS]=K1*[C]+K0r2=0.9991
K0=0.0054,K1=0.7441
Beryllium(Be)
Graphitefurnacetemperatureprogram(Standardmode0.0),Pyrolyticgraphitetube
STAGE#
TEMP.(oC)
TIME(sec)
GAS
(l/min)
HEAT
SAMPLING
PreTIME
1
120
20
#1
1.0
RAMP
OFF
2
250
10
#1
1.0
RAMP
OFF
3
600
10
#1
1.0
STEP
OFF
4
600
3
#1
0.0H
STEP
OFF
5
2600
3
#1
0.0H
STEP
ON
2
CalibrationcurveinBGC-D2mode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)234.9nm
10.00000.0041
20.20000.0412
30.60000.1389
41.00000.2434
[ABS]=K1*[C]+K0r2=0.9973
K0=0.0019K1=0.2416
Bismuth(Bi)
Standardsubstanceandstandardsolutionpreparationmethod
RefertoSection3-3forthestandardsolutionpreparationmethod
Measuringconditionsandcalibrationcurve
Lightingconditions
Current
;10mA/0mA
;10mA/300mA
Wavelength
;223.1nm
;223.1nm
Slitwidth
;
0.5nm
;
0.5nm
Lightingmode
;
BGC-D2
;
BGC-SR
Graphitefurnacetemperatureprogram(Highsensitivitymode0.0H),Highdensitygraphitetube
STAGE#
TEMP.(oC)
TIME(sec)
GAS
(l/min)
HEAT
SAMPLING
PreTIME
1
120
20
#1
1.0
RAMP
OFF
2
250
10
#1
1.0
RAMP
OFF
3
450
10
#1
1.0
STEP
OFF
4
450
3
#1
0.0H
STEP
OFF
5
1700
3
#1
0.0H
STEP
ON
2
CalibrationcurveinBGC-D2mode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)223.1nm
10.00000.0069
210.00000.1326
330.00000.4847
450.00000.7950
[ABS]=K1*[C]+K0r2=0.9981
K0=–0.0065,K1=0.0161
CalibrationcurveinBGC-SRmode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)223.1nm
10.00000.0072
210.00000.0624
330.00000.2143
450.00000.3764
[ABS]=K1*[C]+K0r2=0.9967
K0=0.0030,K1=0.0075
Graphitefurnacetemperatureprogram(Standardmode0.0),Highdensitygraphitetube
STAGE#TEMP.(oC)TIME(sec)GAS(l/min)HEATSAMPLINGPreTIME
1
120
20
#1
1.0
RAMP
OFF
2
250
10
#1
1.0
RAMP
OFF
3
450
10
#1
1.0
STEP
OFF
4
450
3
#1
0.0
STEP
OFF
5
1700
3
#1
0.0
STEP
ON
2
CalibrationcurveinBGC-D2mode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)223.1nm
10.00000.0080
210.00000.0358
330.00000.1334
450.00000.2464
[ABS]=K1*[C]+K0r2=0.9901
K0=–0.0032,K1=0.0048
CalibrationcurveinBGC-SRmode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)223.1nm
10.00000.0071
210.00000.0205
330.00000.0637
450.00000.1107
[ABS]=K1*[C]+K0r2=0.9938
K0=0.0031,K1=0.0021
Ashingtemperature,atomizationtemperatureandsensitivity(Solution:HNO3,Metal)
AshingconditionAtomizationcondition
Calcium(Ca)
Standardsubstanceandstandardsolutionpreparationmethod
RefertoSection3-3forthestandardsolutionpreparationmethod
Measuringconditionsandcalibrationcurve
Lightingconditions
Current
;10mA/0mA
Wavelength
;422.7nm
Slitwidth
;
0.5nm
Lightingmode
;
BGC-D2
Graphitefurnacetemperatureprogram(Highsensitivitymode0.0H),Pyrolyticgraphitetube
STAGE#
TEMP.(oC)
TIME(sec)
GAS
(l/min)
HEAT
SAMPLING
PreTIME
1
120
20
#1
1.0
RAMP
OFF
2
250
10
#1
1.0
RAMP
OFF
3
700
10
#1
1.0
STEP
OFF
4
700
3
#1
0.0H
STEP
OFF
5
2500
3
#1
0.0H
STEP
ON
2
CalibrationcurveinBGC-D2mode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)422.7nm
10.00000.0438
20.40000.2778
31.20000.6807
42.00001.0625
[ABS]=K2*[C]2+K1*[C]+K0r2=0.9998
K0=0.0477,K1=0.5690,K2=–0.0312
Calcium(Ca)
Graphitefurnacetemperatureprogram(Standardmode0.0),Pyrolyticgraphitetube
STAGE#
TEMP.(oC)
TIME(sec)
GAS
(l/min)
HEAT
SAMPLING
PreTIME
1
120
20
#1
1.0
RAMP
OFF
2
250
10
#1
1.0
RAMP
OFF
3
700
10
#1
1.0
STEP
OFF
4
700
3
#1
0.0
STEP
OFF
5
2500
3
#1
0.0
STEP
ON
2
CalibrationcurveinBGC-D2mode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)422.7nm
10.00000.0302
20.40000.2305
31.20000.5795
42.00000.9083
[ABS]=K2*[C]2+K1*[C]+K0r2=0.9999
K0=0.0331,K1=0.4904,K2=–0.0267
Ashingtemperature,atomizationtemperatureandsensitivity(Solution:HCl,CaCO3)
AshingconditionAtomizationcondition
Cadmium(Cd)
Standardsubstanceandstandardsolutionpreparationmethod
RefertoSection3-3forthestandardsolutionpreparationmethod
Measuringconditionsandcalibrationcurve
Lightingconditions
Current
;8mA/0mA
;8mA/300mA
Wavelength
;228.8nm
;228.8nm
Slitwidth
;
0.5nm
;
0.5nm
Lightingmode
;
BGC-D2
;
BGC-SR
Graphitefurnacetemperatureprogram(Highsensitivitymode0.0H),Highdensitygraphitetube
STAGE#
TEMP.(oC)
TIME(sec)
GAS
(l/min)
HEAT
SAMPLING
PreTIME
1
120
20
#1
1.0
RAMP
OFF
2
250
10
#1
1.0
RAMP
OFF
3
300
10
#1
1.0
STEP
OFF
4
300
3
#1
0.0H
STEP
OFF
5
1500
3
#1
0.0H
STEP
ON
2
?CalibrationcurveinBGC-D2mode(Sampleinjectionvolume20l)
STD.
CONC.
ABS.
No.
(ppb)
228.8nm
1
0.0000
0.0129
2
0.1000
0.0751
3
0.3000
0.1971
4
0.5000
0.3340
[ABS]=K1*[C]+K0r2=0.9992
K0=–0.0109,K1=0.6395
CalibrationcurveinBGC-SRmode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)228.8nm
10.00000.0047
20.10000.0363
30.30000.1005
40.50000.1577
[ABS]=K1*[C]+K0r2=0.9993
K0=0.0057,K1=0.3072
Cadmium(Cd)
Graphitefurnacetemperatureprogram(Standardmode0.0),Highdensitygraphitetube
STAGE#
TEMP.(oC)
TIME(sec)
GAS
(l/min)
HEAT
SAMPLING
PreTIME
1
120
20
#1
1.0
RAMP
OFF
2
250
10
#1
1.0
RAMP
OFF
3
300
10
#1
1.0
STEP
OFF
4
300
3
#1
0.0
STEP
OFF
5
1500
3
#1
0.0
STEP
ON
2
CalibrationcurveinBGC-D2mode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)228.8nm
10.00000.0138
20.10000.0453
30.30000.1224
40.50000.1894
[ABS]=K1*[C]+K0r2=0.9989
K0=–0.0127,K1=0.3558
CalibrationcurveinBGC-SRmode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)228.8nm
10.00000.0060
20.10000.0211
30.30000.0568
40.50000.0951
[ABS]=K1*[C]+K0r2=0.9985
K0=0.0045,K1=0.1791
Ashingtemperature,atomizationtemperatureandsensitivity(Solution:HNO3,Metal)
AshingconditionAtomizationcondition
Chromium(Cr)
Standardsubstanceandstandardsolutionpreparationmethod
RefertoSection3-3forthestandardsolutionpreparationmethod
Measuringconditionsandcalibrationcurve
Lightingconditions
Current
;10mA/0mA
;10mA/600mA
Wavelength
;357.9nm
;357.9nm
Slitwidth
;
0.5nm
;
0.5nm
Lightingmode
;
BGC-D2
;
BGC-SR
Graphitefurnacetemperatureprogram(Highsensitivitymode0.0H),Pyrolyticgraphitetube
STAGE#
TEMP.(oC)
TIME(sec)
GAS
(l/min)
HEAT
SAMPLING
PreTIME
1
120
20
#1
1.0
RAMP
OFF
2
250
10
#1
1.0
RAMP
OFF
3
700
10
#1
1.0
STEP
OFF
4
700
3
#1
0.0H
STEP
OFF
5
2500
3
#1
0.0H
STEP
ON
2
CalibrationcurveinBGC-D2mode(Sampleinjectionvolume10l)
STD.CONC.ABS.
No.(ppb)357.9nm
10.00000.0108
21.00000.0807
33.00000.2158
45.00000.3478
[ABS]=K1*[C]+K0r2=0.9999
K0=0.0123,K1=0.0673
CalibrationcurveinBGC-SRmode(Sampleinjectionvolume10l)
STD.CONC.ABS.
No.(ppb)357.9nm
10.00000.0043
21.00000.0444
33.00000.1268
45.00000.2121
[ABS]=K1*[C]+K0r2=0.9999
K0=0.0034,K1=0.0416
Graphitefurnacetemperatureprogram(Standardmode0.0),Pyrolyticgraphitetube
STAGE#TEMP.(oC)TIME(sec)GAS(l/min)HEATSAMPLINGPreTIME
1
120
20
#1
1.0
RAMP
OFF
2
250
10
#1
1.0
RAMP
OFF
3
700
10
#1
1.0
STEP
OFF
4
700
3
#1
0.0
STEP
OFF
5
2500
3
#1
0.0
STEP
ON
2
CalibrationcurveinBGC-D2mode(Sampleinjectionvolume10l)
STD.CONC.ABS.
No.(ppb)357.9nm
10.00000.0124
21.00000.0389
33.00000.1078
45.00000.1661
[ABS]=K1*[C]+K0r2=0.9983
K0=–0.0110,K1=0.0313
CalibrationcurveinBGC-SRmode(Sampleinjectionvolume10l)
STD.CONC.ABS.
No.(ppb)357.9nm
10.00000.0031
21.00000.0236
33.00000.0617
45.00000.1042
[ABS]=K1*[C]+K0r2=0.9994
K0=0.0029,K1=0.0201
Ashingtemperature,atomizationtemperatureandsensitivity(Solution:HCl,Metal)
AshingconditionAtomizationcondition
Cobalt(Co)
Standardsubstanceandstandardsolutionpreparationmethod
RefertoSection3-3forthestandardsolutionpreparationmethod
Measuringconditionsandcalibrationcurve
Lightingconditions
Current
;12mA/0mA
;12mA/400mA
Wavelength
;240.7nm
;240.7nm
Slitwidth
;
0.2nm
;
0.2nm
Lightingmode
;
BGC-D2
;
BGC-SR
Graphitefurnacetemperatureprogram(Highsensitivitymode0.0H),Pyrolyticgraphitetube
STAGE#
TEMP.(oC)
TIME(sec)
GAS
(l/min)
HEAT
SAMPLING
PreTIME
1
120
20
#1
1.0
RAMP
OFF
2
250
10
#1
1.0
RAMP
OFF
3
400
10
#1
1.0
STEP
OFF
4
400
3
#1
0.0H
STEP
OFF
5
2500
3
#1
0.0H
STEP
ON
2
CalibrationcurveinBGC-D2mode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)240.7nm
10.00000.0169
24.00000.1578
312.00000.4332
420.00000.6482
[ABS]=K2*[C]2+K1*[C]+K0r2=0.9997
K0=0.0138,K1=0.0389,K2=–0.0004
CalibrationcurveinBGC-SRmode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)240.7nm
10.00000.0178
24.00000.1058
312.00000.2953
420.00000.4600
[ABS]=K2*[C]2+K1*[C]+K0r2=0.9997
K0=0.0152,K1=0.0243,K2=–0.0000
Cobalt(Co)
Graphitefurnacetemperatureprogram(Standardmode0.0),Pyrolyticgraphitetube
STAGE#
TEMP.(oC)
TIME(sec)
GAS
(l/min)
HEAT
SAMPLING
PreTIME
1
120
20
#1
1.0
RAMP
OFF
2
250
10
#1
1.0
RAMP
OFF
3
400
10
#1
1.0
STEP
OFF
4
400
3
#1
0.0
STEP
OFF
5
2500
3
#1
0.0
STEP
ON
2
CalibrationcurveinBGC-D2mode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)240.7nm
10.00000.0319
24.00000.1063
312.00000.2988
420.00000.4591
[ABS]=K2*[C]2+K1*[C]+K0r2=0.9983
K0=0.0265,K1=0.0227,K2=–0.0000
CalibrationcurveinBGC-SRmode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)240.7nm
10.00000.0154
24.00000.0715
312.00000.2148
420.00000.3384
[ABS]=K2*[C]2+K1*[C]+K0r2=0.9987
K0=0.0117,K1=0.0167,K2=–0.0000
Ashingtemperature,atomizationtemperatureandsensitivity(Solution:HCl,CoCl)
AshingconditionAtomizationcondition
Copper(Cu)
Standardsubstanceandstandardsolutionpreparationmethod
RefertoSection3-3forthestandardsolutionpreparationmethod
Measuringconditionsandcalibrationcurve
Lightingconditions
Current
;6mA/0mA
;10mA/500mA
Wavelength
;324.8nm
;324.8nm
Slitwidth
;
0.5nm
;
0.5nm
Lightingmode
;
BGC-D2
;
BGC-SR
Graphitefurnacetemperatureprogram(Highsensitivitymode0.0H),Pyrolyticgraphitetube
STAGE#
TEMP.(oC)
TIME(sec)
GAS
(l/min)
HEAT
SAMPLING
PreTIME
1
120
20
#1
1.0
RAMP
OFF
2
250
10
#1
1.0
RAMP
OFF
3
500
10
#1
1.0
STEP
OFF
4
500
3
#1
0.0H
STEP
OFF
5
2300
3
#1
0.0H
STEP
ON
2
CalibrationcurveinBGC-D2mode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)324.8nm
10.00000.0041
21.00000.1062
33.00000.3052
45.00000.4881
[ABS]=K1*[C]+K0r2=0.9994
K0=0.0079,K1=0.0969
CalibrationcurveinBGC-SRmode(Sampleinjectionvolume20l)
STD.CONC.ABS.
No.(ppb)324.8nm
10.00000.0025
21.00000.0564
33.00000.1937
45.00000.3153
[ABS]=K1*[C]+K0r2=0.9988
K0=–0.0008,K1=0.0635
Graphitefurnacetemperatureprogram(Standardmode0.0),Pyrolyticgraphitetube
STAGE#TEMP.(oC)TIME(sec)GAS(l/min)HEATSAMPLINGPreTIME
1
120
溫馨提示
- 1. 本站所有資源如無特殊說明,都需要本地電腦安裝OFFICE2007和PDF閱讀器。圖紙軟件為CAD,CAXA,PROE,UG,SolidWorks等.壓縮文件請下載最新的WinRAR軟件解壓。
- 2. 本站的文檔不包含任何第三方提供的附件圖紙等,如果需要附件,請聯(lián)系上傳者。文件的所有權益歸上傳用戶所有。
- 3. 本站RAR壓縮包中若帶圖紙,網(wǎng)頁內容里面會有圖紙預覽,若沒有圖紙預覽就沒有圖紙。
- 4. 未經(jīng)權益所有人同意不得將文件中的內容挪作商業(yè)或盈利用途。
- 5. 人人文庫網(wǎng)僅提供信息存儲空間,僅對用戶上傳內容的表現(xiàn)方式做保護處理,對用戶上傳分享的文檔內容本身不做任何修改或編輯,并不能對任何下載內容負責。
- 6. 下載文件中如有侵權或不適當內容,請與我們聯(lián)系,我們立即糾正。
- 7. 本站不保證下載資源的準確性、安全性和完整性, 同時也不承擔用戶因使用這些下載資源對自己和他人造成任何形式的傷害或損失。
最新文檔
- 2025版學校游泳池兒童游樂區(qū)設計與施工承包合同示范3篇
- 2025版土地使用權出讓居間合同(新型合作模式)3篇
- 2025版城市住宅小區(qū)全面滅蟑螂服務合同4篇
- 2025版土地測繪保密協(xié)議:保密項目合作與技術支持合同3篇
- 乳粉產(chǎn)品質量法律規(guī)制與合規(guī)考核試卷
- 會展產(chǎn)業(yè)與數(shù)字經(jīng)濟的創(chuàng)新結合考核試卷
- 2025版十五年商業(yè)地產(chǎn)租賃合同范本15篇
- 2025版城市慶典活動委托演出合同3篇
- 2025年水土保持設施驗收技術服務與生態(tài)修復實施合同3篇
- 2025年醫(yī)療設備使用及維護管理協(xié)議
- 南通市2025屆高三第一次調研測試(一模)地理試卷(含答案 )
- 2025年上海市閔行區(qū)中考數(shù)學一模試卷
- 2025中國人民保險集團校園招聘高頻重點提升(共500題)附帶答案詳解
- 重癥患者家屬溝通管理制度
- 碳排放管理員 (碳排放核查員) 理論知識考核要素細目表三級
- 2024年河北省中考數(shù)學試題(含答案解析)
- 小學二年級數(shù)學口算練習題1000道
- 納布啡在產(chǎn)科及分娩鎮(zhèn)痛的應用
- DZ/T 0462.4-2023 礦產(chǎn)資源“三率”指標要求 第4部分:銅等12種有色金屬礦產(chǎn)(正式版)
- 化學-福建省龍巖市2024屆高三下學期三月教學質量檢測(一模)試題和答案
- 凸優(yōu)化在經(jīng)濟學與金融學中的應用
評論
0/150
提交評論