1、懸浮物測(cè)定儀使用說(shuō)明書(shū)以下標(biāo)識(shí)將會(huì)在本手冊(cè)或者儀器上出現(xiàn):注 意保險(xiǎn)絲接 地 端目 錄概述 測(cè)量原理································3產(chǎn)品特點(diǎn)·········

2、;·······················3產(chǎn)品應(yīng)用··························

3、······3產(chǎn)品 變送器··································4傳感器········&

4、#183;·························5安裝 注意事項(xiàng)·······················

5、;·········6變送器安裝······························7傳感器安裝·········

6、;·····················8電氣連接····························

7、····11調(diào)試 控制面板································13菜單結(jié)構(gòu)···········

8、83;··················· 14常規(guī)設(shè)置·····························&#

9、183;· 15標(biāo)定····································16維護(hù) 變送器維護(hù)··········

10、;····················18傳感器維護(hù)····························

11、3;·18疑問(wèn) 常見(jiàn)問(wèn)題解答····························19附錄1··················

12、················ 2021一、概 述測(cè)量原理濁度是表現(xiàn)水中懸浮物對(duì)光線(xiàn)透過(guò)時(shí)所發(fā)生的阻礙程度。水中含有泥土、粉塵、細(xì)微有機(jī)物和其他微生物和膠體物可使水中呈現(xiàn)濁度。傳感器上發(fā)射器發(fā)送的光波在傳輸過(guò)程中經(jīng)過(guò)被測(cè)物的吸收、反射和散射后,有一部分透射光線(xiàn)能照射到180°方向的檢測(cè)器上,有一部分散射光散射到90°方向的檢測(cè)器上。在180°和90°方向檢測(cè)器上接收到的光線(xiàn)強(qiáng)度與被測(cè)污水的濁度有

13、一定的關(guān)系,因此通過(guò)測(cè)量透射光和散射光的強(qiáng)度就可以計(jì)算出污水的濁度。產(chǎn)品特點(diǎn)內(nèi)置氣泡消除系統(tǒng)(流通式),提高測(cè)量精度和穩(wěn)定性;清洗維護(hù)非常簡(jiǎn)單,三個(gè)月校正一次;重現(xiàn)性好,不受樣品流速和壓力的影響(投入式);采用兩點(diǎn)校正法; NTU、mg/L兩種測(cè)量單位形式可選。 產(chǎn)品應(yīng)用自來(lái)水廠(chǎng):水庫(kù)、水井、清水池、濾前、濾后、沉淀和出水 廠(chǎng)的濁度監(jiān)測(cè)。工業(yè)過(guò)程: 生產(chǎn)循環(huán)用水、循環(huán)水等水質(zhì)監(jiān)測(cè)。污水處理: 進(jìn)口水、排水口水質(zhì)監(jiān)測(cè)。 二、產(chǎn) 品變送器技術(shù)參數(shù):測(cè)量范圍:流通式,0400NTU 投入式,04000NTU顯示方式:LCD測(cè)量單位:NTU和mg/L兩種可選擇分 辨 率:0.01%精 確 度:

14、7;1.0%FS重 復(fù) 性:±1.0%FS功 率: 10W模擬輸出:420mA、負(fù)載750開(kāi)關(guān)輸出:3個(gè)繼電器、容量220VAC/ 2 A供電電源:AC220V±10%傳感器流通式傳感器技術(shù)參數(shù):材 質(zhì):316不銹鋼線(xiàn) 纜:屏蔽線(xiàn)纜安裝方式:投入式過(guò)程連接:ZG1外形尺寸:60×113mm介質(zhì)壓力:3 bar信號(hào)線(xiàn)纜:10M(標(biāo)配)可延長(zhǎng)防護(hù)等級(jí):IP68工作溫度:傳感器:-2060技術(shù)參數(shù):材 質(zhì):ABS線(xiàn) 纜:屏蔽線(xiàn)纜安裝方式:流通式外形尺寸:280×375mm信號(hào)線(xiàn)纜:10M(標(biāo)配)可延長(zhǎng)防護(hù)等級(jí):IP65工作溫度:傳感器:050 投入式傳感器三、

15、安 裝注意事項(xiàng)儀器應(yīng)選擇安裝在室內(nèi)或有防護(hù)裝置的位置,周?chē)坏梅胖靡兹家妆奈锲?。儀器安裝位置應(yīng)選擇便于用戶(hù)及安裝維護(hù)人員閱讀儀器銘牌,便于使用、維護(hù)及檢修的地方。所有電力和管道連接必須符合國(guó)家和地方標(biāo)準(zhǔn)。儀表電源前端必須安裝絕緣開(kāi)關(guān)或者電路切斷開(kāi)關(guān)。為了安全和避免外部信號(hào)對(duì)儀器造成干擾,儀表電源線(xiàn)應(yīng)接在相應(yīng)規(guī)格、帶有地線(xiàn)標(biāo)志、符合電器標(biāo)準(zhǔn)的插座內(nèi),且地線(xiàn)須確保良好接地。儀器的內(nèi)部電路詳見(jiàn)接線(xiàn)圖,電壓和功率的額定值詳見(jiàn)產(chǎn)品上的銘牌,用戶(hù)電源容量必須滿(mǎn)足正常使用儀器的要求。變送器安裝注意事項(xiàng):避免變送器受陽(yáng)光直射、避免變送器發(fā)生震動(dòng);應(yīng)將變送器安裝在稍高于操作者平視位置,便于操作者瀏覽面板或進(jìn)行

16、控制操作;為變送器箱體的開(kāi)啟和維護(hù)留出足夠的空間。傳感器安裝注意事項(xiàng):將傳感器安裝在工藝的恰當(dāng)位置,以保證獲得具有代表性的測(cè)量結(jié)果。將傳感器安裝在易于觸及的位置,以方便對(duì)傳感器進(jìn)行定期清潔維護(hù)。應(yīng)避免將傳感器安裝在產(chǎn)生氣泡聚集的位置,以防止干擾信號(hào)的產(chǎn)生。將傳感器安裝在有代表性、工藝混合良好的取樣點(diǎn)附近,傳感器和取樣點(diǎn)之間的距離推薦最大值不超過(guò)1.5m(5英尺)。校準(zhǔn)以及正常的使用過(guò)程中,進(jìn)行取樣實(shí)驗(yàn)室分析儀和儀器讀數(shù)結(jié)果進(jìn)行比較是有必要的。不正確的取樣操作是導(dǎo)致測(cè)量數(shù)據(jù)有誤的常見(jiàn)原因。變送器安裝外形尺寸主視圖 側(cè)視圖安裝示意圖安裝緊固變送器背后有三個(gè)孔,尺寸如圖所示,其中上面的孔用于懸掛,下

17、面兩個(gè)孔用于螺栓固定。傳感器安裝支架浸入式安裝(池壁) 1、底座 2、支架托3、防水頭 4、管蓋 5、鎖緊螺釘M6 6、套管 7、安裝管 8、連接套9、傳感器² 固定底座在池壁適當(dāng)位置打入四個(gè)M10鋼制膨脹螺栓。(尺寸見(jiàn)附錄1)² 連接(如左圖所示)將傳感器的電纜依次穿過(guò)連接套、安裝管、套管、管蓋和防水頭;將傳感器的螺紋與連接套的螺紋孔相接并擰緊且需密封處理;不銹鋼連接套則與安裝管相連;² 緊固將安裝管穿過(guò)套管,根據(jù)現(xiàn)場(chǎng)測(cè)量深度,通過(guò)兩個(gè)鎖定螺釘將套管固定在安裝管上的適當(dāng)位置;將管蓋與安裝管擰緊;將防水頭擰緊,固定電纜,防止雨水和其他物質(zhì)進(jìn)入安裝管。²

18、安裝將安裝好的傳感器安裝管的套管裝入安裝在池壁上的支架托上,并用兩個(gè)M6螺栓固定。注意:在連接傳感器與連接管時(shí),請(qǐng)旋轉(zhuǎn)連接管而不要旋轉(zhuǎn)傳感器,否則傳感器的電纜有可能被損壞,同樣在連接安裝管與連接管時(shí)請(qǐng)旋轉(zhuǎn)安裝管。支架浸入式安裝(池頂)1、底座 2、支架托3、防水頭 4、管蓋 5、鎖緊螺釘M6 6、套管 7、安裝管 8、連接套9、傳感器² 固定底座在池壁適當(dāng)位置打入四個(gè)M10鋼制膨脹螺栓。(尺寸見(jiàn)附錄1)² 連接(如左圖所示)將傳感器的電纜依次穿過(guò)連接套、安裝管、套管、管蓋和防水頭;將傳感器的螺紋與連接套的螺紋孔相接并擰緊且需密封處理;不銹鋼連接套則與安裝管相連;²

19、 緊固將安裝管穿過(guò)套管,根據(jù)現(xiàn)場(chǎng)測(cè)量深度,通過(guò)兩個(gè)鎖定螺釘將套管固定在安裝管上的適當(dāng)位置;將管蓋與安裝管擰緊;將防水頭擰緊,固定電纜,防止雨水和其他物質(zhì)進(jìn)入安裝管。² 安裝將安裝好的傳感器安裝管的套管裝入安裝在池壁上的支架托上,并用兩個(gè)M6螺栓固定。注意:在連接傳感器與連接管時(shí),請(qǐng)旋轉(zhuǎn)連接管而不要旋轉(zhuǎn)傳感器,否則傳感器的電纜有可能被損壞,同樣在連接安裝管與連接管時(shí)請(qǐng)旋轉(zhuǎn)安裝管。流通式安裝1、信號(hào)線(xiàn)插座 2、上蓋3、外殼 4、進(jìn)水口 5、出水口 6、排水口 7、溢流口 8、固定孔 ² 在墻壁適當(dāng)位置,根據(jù)固定孔尺寸打入兩個(gè)鋼制螺栓,把控制器懸掛在螺栓上并固定。注:控制器安裝時(shí)

20、不能傾斜,要垂直、水平; 高度要合適,要便于操作使用,維護(hù)保養(yǎng)。² 用帶有雙向接口的水管,連接水箱出水口和傳感器進(jìn)水口。進(jìn)水口前要有控制閥門(mén),以便調(diào)節(jié)進(jìn)水量的大小。流量大約為:1L/min。 ² 在傳感器排水口處安裝排水管將排水引入排水管道。排水管建議用ø50mm以上的硬塑料管,以保證排水順暢。² 用加強(qiáng)型軟塑料管(保證水路暢通),將傳感器出水口引入排水管道。² 檢查以上各步連接是否牢固,有無(wú)滲漏之處,水樣流動(dòng)是否順暢。電氣連接注意事項(xiàng)為了保證工作人員和儀表的安全,請(qǐng)按照如下順序來(lái)完成電氣連接:u 傳感器電纜連接u 420mA、繼電器連接u 2

21、20V交流電纜連接u 確保接線(xiàn)正確后,安裝變送器u 將220V交流電纜與現(xiàn)場(chǎng)220V交流電源相連儀表接線(xiàn)端子圖序號(hào)端子功能備注(1)mA1+4-20mA 輸出正電流輸出 (2)mA1-4-20mA 輸出負(fù)(3)mA2+4-20mA 輸出正備用(4)mA2-4-20mA 輸出負(fù)(5)(6)GND保護(hù)地傳感器屏蔽線(xiàn)(粗黑)(7)VCC電源正傳感器紅線(xiàn)(8)GND電源地傳感器綠線(xiàn)(9)ARS485 總線(xiàn)A傳感器黃線(xiàn)(10)BRS485 總線(xiàn)B傳感器藍(lán)線(xiàn)(11)TXDRS232發(fā)送備用(12)RXDRS232接收(13)NO1繼電器1可設(shè)置的上線(xiàn)報(bào)警(14)COM1繼電器1(15)NO2繼電器2可設(shè)置

22、的下線(xiàn)報(bào)警 (16)COM2繼電器2(17)NO3繼電器3備用(18)COM3繼電器3(19)L交流電源輸入LAC220V(20)N交流電源輸入N(21)PG交流電源保護(hù)地大地四、調(diào) 試控制面板前面板圖及按鍵功能說(shuō)明 MENU鍵,選擇工作模式 SET鍵,執(zhí)行工作和記憶設(shè)定數(shù)值 RUN鍵,進(jìn)入運(yùn)行模式,退出設(shè)置模式 UP鍵,增加數(shù)字 DOWN鍵,減少數(shù)字菜單結(jié)構(gòu)通過(guò)按濁度計(jì)變送器上MODE鍵并輸入正確密碼后就能進(jìn)入菜單界面,菜單是單級(jí)菜單,功能和相關(guān)參數(shù)如表所示。序號(hào)菜單項(xiàng)設(shè)定參數(shù)范圍初始值(流通式)初始值(投入式)1密碼設(shè)置設(shè)定進(jìn)入01000000000002單位設(shè)置設(shè)定顯示單位NTU或 mg

23、/LNTU NTU3小信號(hào)切除設(shè)定小信號(hào)切除值0.0滿(mǎn)量程0.5NTU4量程設(shè)置設(shè)定滿(mǎn)量程濁度05000NTU54mA 校準(zhǔn)校準(zhǔn)4mA 輸出電流值0200010001000620mA 校準(zhǔn)校準(zhǔn)20mA 輸出電流值02000100010007故障電流設(shè)置設(shè)定故障報(bào)警電流值424mA21.0mA21.0mA8濾波系數(shù)設(shè)置設(shè)定測(cè)量值濾波系數(shù)010020209繼電器一開(kāi)通設(shè)置設(shè)定繼電器1 開(kāi)啟濁度值0滿(mǎn)量程根據(jù)要求自行設(shè)定根據(jù)要求自行設(shè)定10繼電器一關(guān)斷設(shè)置設(shè)定繼電器1 關(guān)斷濁度值0滿(mǎn)量程根據(jù)要求自行設(shè)定根據(jù)要求自行設(shè)定11繼電器二開(kāi)通設(shè)置設(shè)定繼電器2 開(kāi)啟濁度值0滿(mǎn)量程根據(jù)要求自行設(shè)定根據(jù)要求自行設(shè)

24、定12繼電器二關(guān)斷設(shè)置設(shè)定繼電器2 關(guān)斷濁度值0滿(mǎn)量程根據(jù)要求自行設(shè)定根據(jù)要求自行設(shè)定13時(shí)間設(shè)置設(shè)定實(shí)時(shí)時(shí)鐘時(shí)間當(dāng)前時(shí)間當(dāng)前時(shí)間14趨勢(shì)線(xiàn)采樣間隔設(shè)定趨勢(shì)線(xiàn)記錄時(shí)間間隔159分鐘30分鐘30分鐘15顯示歷史趨勢(shì)線(xiàn)顯示歷史數(shù)據(jù)趨勢(shì)線(xiàn)16沖洗時(shí)間設(shè)置設(shè)定沖洗時(shí)間1 小時(shí)1 小時(shí)17零點(diǎn)標(biāo)定零點(diǎn)信號(hào)值18零點(diǎn)值修正輸入零點(diǎn)標(biāo)定時(shí)標(biāo)樣實(shí)際濁度值0.0滿(mǎn)量程19標(biāo)準(zhǔn)液標(biāo)定標(biāo)準(zhǔn)液標(biāo)定時(shí)的信號(hào)值20標(biāo)準(zhǔn)液濁度值輸入輸入標(biāo)準(zhǔn)液標(biāo)定時(shí)的濁度值0.0滿(mǎn)量程21標(biāo)定信息標(biāo)定信息22載入出廠(chǎng)標(biāo)定值載入出廠(chǎng)時(shí)的標(biāo)定值常規(guī)設(shè)置 濁度計(jì)安裝完成后,操作人員按照如下順序進(jìn)行設(shè)置后即可正常工作: 繼電器一和繼電器二設(shè)置電流校

25、準(zhǔn)歷史趨勢(shì)線(xiàn)采樣間隔設(shè)置4mA校準(zhǔn)20mA校準(zhǔn)繼電器一開(kāi)通設(shè)置繼電器一關(guān)斷設(shè)置繼電器二開(kāi)通設(shè)置繼電器二關(guān)斷設(shè)置基本設(shè)置單位設(shè)置量程設(shè)置 其它菜單可以使用默認(rèn)的出廠(chǎng)設(shè)置,也可以根據(jù)實(shí)際情況進(jìn)行相應(yīng)設(shè)置。標(biāo)定零點(diǎn)標(biāo)定:* 17. 零點(diǎn)標(biāo)定18. 零點(diǎn)值修正19. 標(biāo)準(zhǔn)液標(biāo)定20. 標(biāo)準(zhǔn)液濁度值輸入零點(diǎn)標(biāo)定00000mV SET 確定 RUN 取消零點(diǎn)修正值:17. 零點(diǎn)標(biāo)定* 18. 零點(diǎn)值修正19. 標(biāo)準(zhǔn)液標(biāo)定20. 標(biāo)準(zhǔn)液濁度值輸入零點(diǎn)值修正(1021) 000.0 NTU標(biāo)準(zhǔn)液標(biāo)定:17. 零點(diǎn)標(biāo)定18. 零點(diǎn)值修正* 19. 標(biāo)準(zhǔn)液標(biāo)定20. 標(biāo)準(zhǔn)液濁度值輸入標(biāo)準(zhǔn)液標(biāo)定(1117) 100

26、0.0mVSET 確定 RUN 取消標(biāo)準(zhǔn)液濁度值輸入:17. 零點(diǎn)標(biāo)定18. 零點(diǎn)值修正19. 標(biāo)定液標(biāo)定*20. 標(biāo)準(zhǔn)液濁度值輸入標(biāo)準(zhǔn)液濁度值輸入(3800) 1200NTU標(biāo)準(zhǔn)液配制: 儀器濁度標(biāo)準(zhǔn),執(zhí)行GB5750-85生活飲用水標(biāo)準(zhǔn)檢驗(yàn)法,第6.2.4條和生活飲用水衛(wèi)生規(guī)范第5.1.3條。零點(diǎn)標(biāo)定在出廠(chǎng)時(shí)已經(jīng)標(biāo)定過(guò),用戶(hù)只有在使用時(shí)間較長(zhǎng)需要校準(zhǔn)時(shí)才需要再次標(biāo)定,零點(diǎn)標(biāo)定的步驟如下:u 將傳感器清洗干凈,去除表面污垢;u 將傳感器置于盛有清水的容器中,使用深顏色的容器(黑色),避免陽(yáng)光直射;u 儀表在清水中穩(wěn)定十五分鐘后,進(jìn)入“零點(diǎn)標(biāo)定”界面;u 確認(rèn)操作正確,無(wú)陽(yáng)光直射、傳感器清洗干

27、凈等條件后。按“SET”鍵進(jìn)行確定,此時(shí)儀表已經(jīng)記錄了清水中的值。如誤操作請(qǐng)按RUN鍵退出菜單。u 正常情況零點(diǎn)修正值不需要修改,如果需要修改請(qǐng)調(diào)至“18.零點(diǎn)值修正”處。u 按“SET”鍵進(jìn)入“零點(diǎn)值修正”界面,通過(guò)“” 和 “”鍵輸入零點(diǎn)修正值,按“SET”鍵保存數(shù)據(jù)并退出“零點(diǎn)值修正”界面。u 將傳感器清洗干凈,確保光束的發(fā)送與接收窗口沒(méi)有雜物遮攔;u 將傳感器置于盛有標(biāo)準(zhǔn)液(詳見(jiàn)標(biāo)準(zhǔn)液配制)的容器內(nèi),儀表正常工作十五分鐘后,進(jìn)入“19. 標(biāo)準(zhǔn)液標(biāo)定”處;觀(guān)察標(biāo)準(zhǔn)液的標(biāo)定值是否正常,如果標(biāo)定值正常則按“SET”鍵保存數(shù)據(jù),如果標(biāo)定值不正常請(qǐng)按“RUN”鍵退出菜單并重新開(kāi)始標(biāo)準(zhǔn)液標(biāo)定。u

28、進(jìn)入“標(biāo)準(zhǔn)液濁度值輸入”界面,將標(biāo)準(zhǔn)液的濁度值輸入,按“SET”鍵保存數(shù)據(jù);例如化驗(yàn)結(jié)果為1200NTU,在“標(biāo)準(zhǔn)液濁度值輸入”界面通過(guò)“”和 “” 鍵將濁度值調(diào)整到1200NTU,然后按SET確定,通過(guò)“”和 “” 輸入已測(cè)得的濁度值,按“SET”鍵保存數(shù)據(jù)并退出修正界面,完成標(biāo)準(zhǔn)液濁度值輸入。配置方法如下:1、配置0 NTU純水取蒸餾水經(jīng)過(guò)0.2um膜濾器過(guò)濾。2、硫酸肼溶液(10g/L)稱(chēng)取硫酸肼(又名硫酸聯(lián)氨) 10.0g溶于純水,于100mL容量瓶中定容。3、六甲基四氨溶液(100g/L)稱(chēng)取六甲基四氨10.0溶于純水,于100mL容量瓶 中定容。4、福爾馬肼濁度標(biāo)準(zhǔn)液分別吸取硫酸肼

29、溶液5.00mL、六甲基四氨溶液5.00mL于100ml容量瓶?jī)?nèi),混勻,在25±3放置24小時(shí)后,加純水至刻度,混勻。此標(biāo)準(zhǔn)液濁度為400(NTU)。5、100 NTU福爾馬肼濁度標(biāo)準(zhǔn)液吸取400NTU福爾馬肼標(biāo)準(zhǔn)混合液250mL置于1000ml容量瓶?jī)?nèi),加純水至刻度,混勻。此標(biāo)準(zhǔn)液為100NTU。五、維 護(hù)變送器的維護(hù)傳感器的維護(hù)變送器根據(jù)使用的要求,安裝位置和工作情況比較復(fù)雜,為了使變送器正常工作維護(hù)人員需要對(duì)變送器進(jìn)行定期維護(hù),維護(hù)時(shí)請(qǐng)注意如下事項(xiàng):u 安裝在室外請(qǐng)檢查變送器安裝箱體是否有漏水等現(xiàn)象;u 檢查變送器的工作環(huán)境,如果溫度超出變送器的工作穩(wěn)定范圍,請(qǐng)采取相應(yīng)措施,否則變送器可能損壞或降低使用壽命;u

溫馨提示

  • 1. 本站所有資源如無(wú)特殊說(shuō)明,都需要本地電腦安裝OFFICE2007和PDF閱讀器。圖紙軟件為CAD,CAXA,PROE,UG,SolidWorks等.壓縮文件請(qǐng)下載最新的WinRAR軟件解壓。
  • 2. 本站的文檔不包含任何第三方提供的附件圖紙等,如果需要附件,請(qǐng)聯(lián)系上傳者。文件的所有權(quán)益歸上傳用戶(hù)所有。
  • 3. 本站RAR壓縮包中若帶圖紙,網(wǎng)頁(yè)內(nèi)容里面會(huì)有圖紙預(yù)覽,若沒(méi)有圖紙預(yù)覽就沒(méi)有圖紙。
  • 4. 未經(jīng)權(quán)益所有人同意不得將文件中的內(nèi)容挪作商業(yè)或盈利用途。
  • 5. 人人文庫(kù)網(wǎng)僅提供信息存儲(chǔ)空間,僅對(duì)用戶(hù)上傳內(nèi)容的表現(xiàn)方式做保護(hù)處理,對(duì)用戶(hù)上傳分享的文檔內(nèi)容本身不做任何修改或編輯,并不能對(duì)任何下載內(nèi)容負(fù)責(zé)。
  • 6. 下載文件中如有侵權(quán)或不適當(dāng)內(nèi)容,請(qǐng)與我們聯(lián)系,我們立即糾正。
  • 7. 本站不保證下載資源的準(zhǔn)確性、安全性和完整性, 同時(shí)也不承擔(dān)用戶(hù)因使用這些下載資源對(duì)自己和他人造成任何形式的傷害或損失。

最新文檔

評(píng)論

0/150

提交評(píng)論